The Athos series patient monitor are designed to meet the daily clinical needs about the sub-intensive patient monitoring.
Features
Multiple screen layout to view big font size in numeric and waveform.
Trending facilities for up to 360 hours of both graphical & numerical along with event review facility for all parameters and full waveform disclosure facility.
Different visual alarm with colour coding and audible alarms for various parameters with three levels of volume adjustment.
Support HL7 protocol to access the central workstation Energy-saving, low-power design, extending battery life Configure noise reduction fan to ensure a quieter and cleaner environment Arrhythmia detection and analysis in various leads.
A inbuilt battery backup of 4 hours.
Monitor should be modular in design, compact and suitable for adult, pediatric and neonates.
Standard parameters: SpO2, 12leads ECG, PR, HR, RESP, NIBP,2*TEMP.
Optional parameters: EtCO2, 2*IBP, CO.
12.1inches/15.6inches colour LED screen with touch screen and /or rotator knob.
Upgradable module is interchangeable (plug and play type).
Capable to visualize 6 or more waveform.
Support colour coding for different waveform.
ST segment analysis with ST trend with continuous display of ST value on screen
PERFORMANCES
SPO2
SPO2 Range : 0 ~ 100 %
SPO2 Accuracy : 0 ~ 100 %, ± 2% <70 %, Undefined
PI Range : 0 ⁓ 20 %
PVI Range : 0.001 %
PR Range : 25 – 250 bpm
PR Accuracy : ± 2 bpm or ± 2 % (whichever is greater)
ECG
ECG Range : 0.15 ⁓ 5.5 mV
ECG Resolution : 2.36 uV/LSB
HR Range : 15 ⁓ 300 bmp (adult) 15 ~ 250 bpm (child/neonate)
HR Accuracy : ± 1 bpm or ± 1 % (whichever is greater)
RR Range : 0 ~ 120 bpm
RR Accuracy : 15 ~ 120 rpm : ± 2 rpm or ± 2% <15 rpm : Undefined
NIBP
Measurement Range :
Systolic : Adult mode : 40 ~ 270 mmHg, Pediatric mode: 40 ~ 200 mmHg, Neaonate : 40 ~ 130 mmHg
Diastolic : Adult mode : 10 ~ 210 mmHg, Pediatric mode: 10 ~ 162 mmHg, Neaonate : 10 ~ 90 mmHg
Mean : Adult mode: 20 ~ 230 mmHg, Pediatric mode: 20 ~ 170 mmHg, Neaonate : 20 ~ 100 mmHg
Accuracy : Mean error : < 5mmHg, Standard deviation : < 8 mmHg
Pressure Range : 0 mmHg (0kPa) ~ 300mmHg (39.9kPa)
Pressure Accuracy : ±2 mmHg or ±1% of reading
Resolution : 1 mmHg
Temp
Range : 0 – 50 °C
Accuracy : ± 0.1°C
Resolution : 0.1 °C
RESP
Range : 0 – 120 Rpm
Accuracy : 15 – 120 Rpm (±2 Rpm or ±2% of the reading (whichever is larger))
Resolution : 1 Rpm
ENVIRONMENTAL CHARACTERISTICS
Operation
Temperature: -0℃ ~ 40℃
Humidity: 10% ~ 95%, non-condensing
Atmospheric pressure: 70.0kPa ~ 106.0kPa
Storage
Temperature: -20℃ ~ 60℃
Humidity: 10% ~ 95%, non-condensing
Atmospheric pressure: 57.3kPa ~ 106.0kPa
Downloads