The Porthos series patient monitor are designed to meet the daily clinical needs about the sub-intensive patient monitoring.
Features
LowPulseStrTM SPO2 algorithm more reliable readings of SPO2 during low perfusion and motion
iFastBP NIBP algorithm used for fast and comfortable measurement
wSmartHeart ECG technology for more safety monitoring of patient
iRealRespTM Breating rate technology to get the real reading during motion
wSmartGas Capnograph technology to get reliable reading during multi-environment
Multi display mode other beds view/Big font
Wifi connectivity more flexible application
10,12,15inch screen can be optional
Easy to use with excellent usability
eTeleView Central Montior system support
Smart touch screen for easy operation
PERFORMANCES
SPO2
SPO2 Range : 0 ~ 100 %
SPO2 Accuracy : 0 ~ 100 %, ± 2% <70 %, Undefined
PI Range : 0 ⁓ 20 %
PVI Range : 0.001 %
PR Range : 25 – 250 bpm
PR Accuracy : ± 2 bpm or ± 2 % (whichever is greater)
ECG
ECG Range : 0.15 ⁓ 5.5 mV
ECG Resolution : 2.36 uV/LSB
HR Range : 15 ⁓ 300 bmp (adult) 15 ~ 350 bpm (child/neonate)
HR Accuracy : ± 1 bpm or ± 1 % (whichever is greater)
RR Range : 0 ~ 120 bpm
RR Accuracy : 15 ~ 120 rpm : ± 2 rpm or ± 2% <15 rpm : Undefined
NIBP
Measurement Range :
Systolic : Adult mode : 40 ~ 270 mmHg, Pediatric mode: 40 ~ 200 mmHg, Neaonate : 40 ~ 130 mmHg
Diastolic : Adult mode : 10 ~ 210 mmHg, Pediatric mode: 10 ~ 162 mmHg, Neaonate : 10 ~ 90 mmHg
Mean : Adult mode: 20 ~ 230 mmHg, Pediatric mode: 20 ~ 170 mmHg, Neaonate : 20 ~ 100 mmHg
Accuracy : Mean error : < 5mmHg, Standard deviation : < 8 mmHg
Pressure Range : 0 mmHg (0kPa) ~ 300mmHg (39.9kPa)
Pressure Accuracy : ±2 mmHg or ±1% of reading
Resolution : 1 mmHg
Temp
Range : 0 – 50 °C
Accuracy : ± 0.1°C
Resolution : 0.1 °C
RESP
Range : 0 – 120 Rpm
Accuracy : 15 – 120 Rpm (±2 Rpm or ±2% of the reading (whichever is larger))
Resolution : 1 Rpm
ENVIRONMENTAL CHARACTERISTICS
Operation
Temperature: -0℃ ~ 40℃
Humidity: 10% ~ 95%, non-condensing
Atmospheric pressure: 70.0kPa ~ 106.0kPa
Storage
Temperature: -20℃ ~ 60℃
Humidity: 10% ~ 95%, non-condensing
Atmospheric pressure: 57.3kPa ~ 106.0kPa
Downloads
Click to download