SpotLife® Touch Vital signs monitor has been designed to serve frontline caregivers in emergency, perioperative care and ICU/NICU departments. With its accurate oximetry adapted for adults as well as neonate, it provides a continuous and accurate monitoring of the SpO2 and PR even in case of low perfusion. SpotLife® VSM is available in versions with either NIBP or CO2 measurement with optional fast temperature monitoring. Its handle and lightweight make it easy for transport use while its 4 stable feet allow a reliable use as bedside monitor. Cost effective, it remains a comprehensive device for multiple applications and environments. Equipped of a touch screen and a knob it is extremely easy and smooth to navigate through the menus and selection. Its 3 backlighted buttons for alarm, start/stop NIBP measurement and power are visible at anytime in any light condition.
Features
It is equipped with a 5′ TFT display and the touch activated display panel is inclined at 15°, which is convenient for doctors to view.
72-hour data storage , uninterrupted recording trend data, with dual alarm function of sound and light , alarm parameters
RS232 serial port data transmission function.
Multiple models, standard configuration SPO2,
List of possible configurations:
SPLF-OXT: SpO2, PR, PI
SPLF-NPT: SPO2, PR, PI, NIBP
SPLF-MCT: SpO2, PR, PI, Mainstream CO2 with EtCO2, FiCO2, AwRR
SPLF-SCT: SpO2, PR, PI, Internal Sidestream CO2 with EtCO2, FiCO2, AwRR
SPLF-SET: SpO2, PR, PI, External Sidestream CO2 with EtCO2, FiCO2, AwRR
SPLF-SWT: SPO2, PR, PI, Internal Watertrap Sidestream CO2 with EtCO2, FiCO2, AwRR
All models above can have optional TEMP measurement with IR Non Touch temperature measurements
ENVIRONMENTAL CHARACTERISTICS
Operation
Temperature: -0℃ ~ 40℃
Humidity: 10% ~ 95%, non-condensing
Atmospheric pressure: 70.0kPa ~ 106.0kPa
Storage
Temperature: -20℃ ~ 60℃
Humidity: 10% ~ 95%, non-condensing
Atmospheric pressure: 57.3kPa ~ 106.0kPa
ELECTRICAL AND MECHANICAL CHARACTERISTICS
Electrical Power Supply
AC: 100 ~ 240V ± 10% (built-in adapter)
Input power: 60VA
Frequency: 50Hz/60Hz ± 3Hz
One equipotential grounding interface
Product Class: IIb (in accordance with MDD93 /42/ EEC)
Battery
Built-in rechargeable lithium batteries: 11.1 V / 2400mAh
Continuous runtime on battery: 4 ~ 5 hours depending on use. Optional runtime up to 8h with SpotLife Touch SpO2 model only (built-in larger battery, to be requested on order).
Mechanical
Dimensions: 255*140*95mm (LxWxH)
Weight: 1.5 kg (without accessories)
INTERFACES
Connectlvlty
USB interface
RS232 interface
Connected to central monitor via RJ45
Bluetooth Printer
Ethernet Port
HMI
Optional : 6 models of configuration options
Display: 5” Color TFT LCD, 800 x 480pixels
Audio/Visual Indicators: Alarm limit reached, Alarm tone, Alarm mute, pulse
strength, Patient name, Patient Type, Time, battery status, connection status.Display Indicators: Patient type, Pressure Unit or SpO2/PR (according to model), battery in use, battery charging, silenced alarm.
User Interface language: English (additional language upon request)
PERFORMANCES
Oximetry(All models):
SPO2: Measurement Range: 0 – 100%, Accuracy Range: 70 – 100 %, ±2% < 70%, Undefined, Resolution: 1%
Pulse Rate: Measurement Range: 15-300 bpm, Accuracy Range: ±2%, Resolution: 1 bpm
PI: Measurement Range: 0-20%, Resolution: 0.001%
PIV: Measurement Range: 0-100%, Accuracy Range: Undefined, Resolution: 1%
Ethernet Port
NIBP (SPLF-NP model only)
METHOD: Automatic oscillometric
OPERATION MODES: Manual, Automatic, Continuous
AUTOMATIC MODE MEASUREMENT INTERVAL: 1min/2min/3min/4min/5min/10min/15min/30min/60min/90min/2h/3h
CONTINUOUS MODE MEASUREMENT PERIOD: 5mins, with 5s between each measurement
MAXIMUM SINGLE MEASUREMENT TIME: 180s
MEASUREMENT RANGE: Systolic BP: Adult mode: 40 ~ 270 mmHg, Pediatric mode: 40 ~ 200 mmHg,
Diastolic BP: Adult mode: 10 ~ 210 mmHg, Pediatric mode: 10 ~ 162 mmHg,
MAP: Adult mode: 20 ~ 230 mmHg, Pediatric mode: 20 ~ 175 mmHgACCURACY: Mean error: <±5mmHg, Standard deviation: <8mmHg
STATIC PRESSURE MEASUREMENT RANGE: 0 mmHg (0kPa) ~ 300mmHg (39.9kPa)
STATIC PRESSURE MEASUREMENT ACCURACY: ±2mmHg or ±1% of reading (Whichever is greater)
RESOLUTION: 1 mmHg
INITIAL INFLATION PRESSURE SETTING RANGE: Adult mode: 80 ~ 280 mmHg, Pediatric mode: 80 ~ 210 mmHg
INITIAL INFLATION PRESSURE DEFAULT: Adult mode: 160 mmHg, Pediatric mode: 140 mmHg
SOFTWARE OVER-PRESSURE PROTECTION: Adult mode: 297 ±3mmHg, Pediatric mode: 240 ±3mmHg
ALARM RANGE: Systolic BP: Adult mode: 40 ~ 270 mmHg, Pediatric mode: 40 ~ 200 mmHg,
Diastolic BP: Adult mode: 10 ~ 210 mmHg, Pediatric mode: 10 ~ 162 mmHg,
MAP: Adult mode: 20 ~ 230 mmHg, Pediatric mode: 20 ~ 175 mmHg
Capnography
METHOD: Infrared radiation absorption technology
CO2 MEASUREMENT RANGE: 0 ~ 20 Vol%
ACCURACY: 0 ~ 12 %: ± (0.2 Vol% + 2% of reading), 12 ~ 20 % : ± (0.2 Vol% + 6% of reading)
MEASUREMENT ACCURACY DRIFT: accuracy requirements within 6 hours
RESOLUTION: 0.1 Vol%
ACCURACY: Mean error: <±5mmHg, Standard deviation: <8mmHg
APNEA ALARM DELAY TIME: 20s, 25s, 30s, 35s, 40s, 45s, 50s, 55s, 60s
ALARM RANGE: EtCO2: 0 ~ 150mmHg, FiCO2: 0 ~ 150mmHg, awRR: 0 ~ 150rpm
Thermometry
METHOD: Infrared non-contact forefront measurement
TEMP MEASUREMENT RANGE: 34 ~ 42.9 °C
ACCURACY: 34.0 ~ 34.9 °C ± 0.3 °C, 35.0 ~ 42.0 °C ± 0.2 °C, 42.1 ~ 42.9 °C ± 0.3 °C
RESOLUTION: 0.1 °C
ALARM RANGE: 34 ~ 42 °C
ORDERING
Downloads
Click to download