SpotLife® Veterinary is a vital signs monitor that comes in 3 versions for the benefit of the veterinary market. The SpO2-Only model helps accurate oximetry monitoring of even the smallest animals thanks to its high performances in low perfusion condition. The 2 other models include capnography (mainstream or sidestream) along with oximetry. They will appear to be indispensable and reliable for intubated animals during surgery.
The clear and easy-to-read LCD will show real-time pleth and capnogram, allowing vets to confirm proper intubation, and whether the animal is in respiratory distress. SpotLife compatible airway adapters allow a smooth connection to any anesthesia systems.
Its handle and light weight make it easy for transport use as well as use in operating theaters. Cost effective, it remains a comprehensive device for multiple applications on a wide range of animals and environments.
All models are based on our SpO2 module and provide the same high performance, while our CO2 models will be using either our CO2 mainstream or external sidestream modules.
SPECIFICATIONS
Functions
Oximetry (all models)
Real-time measurement of SpO2, pulse rate (PR), perfusion index (PI), and Variability of Pulse Index (PIV).
Real-time photoplethysmogram display
High accuracy and reliability even with low blood perfusion down to 0.05%, allowing application on small animals.
Capnography (SPLFV-SE and SPLFV-MC models)
Measurement of EtCO2, FiCO2 and Respiratory Rate (awRR)
Real-time capnogram display
Automatic barometric pressure, temperature compensation and manual anaesthetic gases compensation
Fixed Micro Flow: 50 ml/min (±10ml/min) (SPLFV-SE Model only)
Thermometry (Option TBD):
Body temperature measurement using Fast temperature probe
Environmental characteristics
Operating Conditions
Temperature: 0°C ~ 40°C
Humidity: 15 ~ 95%, non-condensation
Atmospheric pressure: 70.0kPa ~ 106.0kPa
Storage Conditions
Temperature: -20 ~ 70°C (4 to 158°F)
Humidity: 15 ~ 95%, non-condensation
Atmospheric pressure: 57.3kPa ~ 106.0kPa
Electrical and Mechanical characteristics
Electrical Power Supply
AC: 100 ~ 240V ± 10% (built-in adapter)
Input power: 60VA
Frequency: 50Hz/60Hz ± 3Hz
One equipotential grounding interface
Product Class: IIb (in accordance with MDD93 /42/ EEC)
Battery
Built-in rechargeable lithium batteries: 11.1 V / 2400mAh
Mechanical
Dimensions: 255 x 140 x 95mm (LxWxH)
Weight: < 2kg (without accessories)
Interfaces
Connectivity
USB interface
RS232 DB9 interface
HMI
Five buttons: Power switch, alarm stop, alarm pause, up and down buttons, menu
Display: 5” Color TFT LCD, 800 x 480pixels
Audio/Visual Indicators: Alarm limit reached, Alarm tone, Alarm mute, pulse strength, Patient name, Patient Type, Time, battery status, connection status.
LEDs: Adult, Neonate, SpO2/PR (according to model), battery in use, battery charging, silenced alarm.
User Interface language: English, French, Italian, Hungarian (additional language upon request).
Performances
SpO2 (all models)
Sp02: Measurement Range: 0 – 100%; Accuracy Range: from 70 – 100 % = ±2%, below 70%, Undefined; Resolution: 1%
Pulse Rate: Measurement Range: 25-250 BPM; Accuracy Range: ±2%; Resolution: 1 bpm
PI: Measurement Range: 0-20%; Resolution: 0.001%
PIV: Measurement Range: 0-100%; Accuracy Range: Undefined; Resolution: 1%
Capnography (SPLFV-MC and SPLFV-SE models)
Method: Infrared radiation absorption technology
CO2 measurement range: 0 ~ 20 %
Accuracy:
0 ~ 12% : ±(0.2 Vol% + 2% of reading)
12 ~ 20% : ±(0.2 Vol% + 6% of reading)
Measurement accuracy drift: accuracy requirements within 6 hours
Resolution: 0.1 Vol%
Accuracy: Mean error: < ±5mmHg, Standard deviation: < 8mmHg
Apnea alarm delay time: 20s, 25s, 30s, 35s, 40s, 45s, 50s, 55s, 60s
Alarm range
EtCO2: 0 ~ 150mmHg
FiCO2: 0 ~ 150mmHg
AwRR: 0 ~ 150rpm
Features
Bright, large 5” Color TFT LCD for better readability of all parameters.
Intuitive user interface.
3 models availables:
SpO2 only (Model #SPLFV-OX)
SpO2 + Mainstream CO2 (Model #SPLFV-MC)
SpO2 + Sidestream CO2 (Model #SPLFV-SE)
Selectable Display Modes:
digits only, showing parameters values
digits and graphs, showing real-time plethysmogram and capnogram (according to model)
digits, graph, and trends, displaying, in addition, the historical values.
9 LED indicators to easily control patient type, units, power and alarm status.
New patient presets allowing a one-step action to set all alarms limits according to patient type (Adult, Child, Neonate).
360° Visible and Audible alarms.
Over 72h trend data with a 1 min resolution.
Built-in power supply and 2400mAh lithium battery for 6h continuous monitoring.
Extension port upon request for custom protocols, raw data output for research and other device connectivity.
USB port for easy updates with new software functions (further connectivity and control options, addition of peripherals).
Ordering Information
Model #SPLFV-OX : SpotLife® Oximeter Only, with reusable animal tongue/ear clip SpO2 probe
Model #SPLFV-MC : SpotLife® Oximeter and Mainstream CO2 sensor, with reusable animal tongue/ear clip SpO2 probe, one set of disposable small and big airway adapters for small and large animal applications
Model #SPLFV-SE : SpotLife® Oximeter and External Sidestream CO2 sensor, with reusable animal tongue/ear clip SpO2 probe, CO2 microstream sampling line.
Downloads
Click to download