Our capnography technology is available in inexpensive and reliable mainstream or sidestream versions for manufacturers of patient monitors, anaesthesia machines and ventilators.
SensoGas-SE external sidestream capnograph module (#SGSE) is an excellent plug and play alternative module to added to your designs for your CO2 gas analysis needs. It can replace seamlessly competing sidestream capnograph modules on the market, consult us for details.
It is used in conjonction with our new PFSA polymer sampling line that takes off the humidity allowing long working hours without replacement of the sampling line. Pump lifetime is prolongated thanks to the detection of the sampling line presence. When the sampling line is not present, the pump would simply go in standby.
You may also be interested in our mainstream capnograph module for your anaesthesia, ventilators or emergency applications.
Features
Provides real-time data of EtCO2, InsCO2, Capnogram and Respiration Rate
Supports monitoring patients with high respiration rates, up to 150 rpm
Supports intubated and non-intubated patients
Sensitive CO2 detection, can detect minimum 0.1 vol% CO2 change
High accuracy with microflow sampling at 50ml/min
Low rise-up Time (< 180ms), Response Time below 2s
Long continuous working time (depending on model)
Wide range of gas measurement (0 ~ 20%)
Sensors for Software and Hardware status
Capnogram data
Automatic zeroing
Microstream sampling line made of PFSA polymer to remove humidity
Sampling line presence detection to extend pump lifetime
Simple interfacing with RS232 serial interface via DB9 connector
Accuracy according to the requirements of ISO 80601-2-55:2011
For OEM version, can be delivered with compatible communication protocols
SPECIFICATIONS
Electrical and Mechanical Specifications
Voltage: +5V DC, offset voltage between ±10% of full range
Power: Normal operation ≤ 1.5W
Dimension: 90mm×60mm×31mm
Weight: ≤ 180g (excluding cable)
Performances, Range and Accuracy
CO2
Range: 0 ~ 20 vol %
Resolution: 0.1 vol %
Accuracy: from 0 ~ 12 vol %, ±(0.2vol% + 2% of reading);
from 12 ~ 20 vol %, ±(0.2vol% + 6% of reading)
Respiratory Rate
Range: 0 ~ 150rpm
Accuracy: from 0 ~ 70rmp, ±1rpm; above 70rmp, undefined
Resolution: ±1rpm
Timings
Rise time: ≤ 180ms
System response time: ≤ 2s (with a 2m sampling line)
Fixed Micro Flow: 50 ml/min (±10ml/min)
Ratings
Working Conditions
Temperature: 10 ~ 40°C (50 to 104°F)
Humidity: 15 ~ 90%, non-condensation
Altitude: 500 ~ 1200hPa
Storage Conditions
Temperature: -20 ~ 70°C (4 to 158°F)
Humidity: 15 ~ 95%, non-condensation
Altitude: 400 ~ 1200hPa
Downloads
Click to download