12 Leads Electrocardiograph Module - MECG12
Our 12 Leads ECG module is suitable for emergency monitoring, often integrated in Patient Monitors. It provides high performances for the diagnostic of arrhythmia, conduction block, myocardial infarction and chamber room hypertrophy. Its analysis capabilities comply with the MIT database.
This module has been integrated in several types of ECG systems designed by our clients, which are CE certified and already sold worldwide. It provides the electrical isolation and the performances required to pass all certification tests, ensuring our customers a successful launch of their products.
It is suitable for integration in patient monitors.
Features
Provides 12 leads ECG monitoring
4 measurement modes : diagnose, monitor, HARDEST and operation measurement
Special function of ECG analysis, verified by MIT database.
Arrhythmia diagnosis.
Patient electrical isolation for safety
Compliant with IEC60601-2-25 standard
Compatible with ZugMed protocol
For integration in Patient monitors
Customised design and protocol upon request.
SPECIFICATIONS
Electrical and Mechanical Specifications
Voltage: +12V DC, offset voltage between ±10% of full range
Power consumption: ≤ 3W
Dimension: 130mm×70mm×13mm
Performances
Input impedance: 5MΩ
Common mode rejection ratio: diagnose mode 100dB, monitor mode 105dB
Polarization voltage: ±450mV
Baseline recovery time: < 3s
Noise: < 30μV
Wave Filtering Mode:
Low-pass filtering: 70Hz, 100Hz, 150Hz
High-pass filtering: 0.05Hz, 0.1Hz, 0.2Hz, 0.5Hz
Myoelectricity filtering: 25Hz, 35Hz, 45Hz
Alternative current filtering: 50Hz, 60Hz, None
Average false detection rate: 0.26%
Average sensitivity: 99.8%
Average positive predictive rate: 99.94% (MIT database)
Range and Accuracy
Pulse Rate
Range: 30 ~ 300bpm
Accuracy: ±1%
Respiration
Range: 15 ~ 120rpm
Accuracy: ±1rpm
Temperature
Range: -1°C ~ 60°C
Accuracy: ±0.1% from 25°C ~ 45°C; otherwise ±0.2% between -1°C ~ 60°C
Downloads
Click to download