4-Channels Invasive Blood Pressure Module
Our 4-Channel IBP module provides the measurements of Systolic and Diastolic Blood Pressure, Mean Arterial Pressure as well as Pulse Rate. The response time is configurable and it is simple to integrate with the provided source code to speed up your time to market.
It complies with IEC 60601-2-34:2000, and has already been integrated in CE marked devices.
For more hemodynamics parameters, you may also be interested in our Continuous CO and 2Ch IBP module for patient monitoring.
Features
Small size for simple assembly
4 channels IBP monitoring of SBP, DBP, MAP and PR
Response time settings
Compatible with ZUGMED protocol
Compliance with 60601-2-34
Easy integration with source code and support provided
SPECIFICATIONS
Electrical and Mechanical Specifications
Voltage: +12V DC, ±0.5V
Power Consumption : < 1.5W
Dimension: 87mm×45mm×15mm
Performances, Range and Accuracy
Blood Pressure
Range: -50 ~ 400mmHg
Accuracy: ±2mmHg, or ±1% of reading (whichever is bigger)
Pulse Rate
Range: 35~250bpm
Accuracy: ±3bpm