MMP12 Multi-parameters Module
Application
SpO2, NIBP, 12 Leads ECG, Temperature, Respiration Rate, PR functions of patient monitor and emergency transport monitors.
Functions
ECG measurement
The result of ECG measurement maximum support 12leads ECG waveform includes heart rate, body temperature, respiration, and ST-segment deviation value of channel I, II and V1 as well as the arrhythmia.
Four measurement modes are provided: diagnosis, monitoring, HARDEST and operation.
Three patient models are supported: adult, pediatric and neonate.
Calibration for ECG: input 1mv standard voltage to record waveform amplitude.
Channel of Heart rate calculation/Arrhythmia analysis: Heart Rate I, II and V1 as well as self-select are for selection.
Notch Mode: 50 Hz, 60 Hz, and 50/60 Hz and off are for selection.
Interval measurement of ECG waveform characteristic are supported: involving measurement of PR interval, QRS interval, QT interval, QTC, P axis, QRS electrical axis and T axis as well as amplitude measurement of RV5, SV1.
Printable diagnostic analysis are provided including arrhythmia, conduction block, myocardial infarction, STT variations, ventricular hypertrophy, atrial enlargement, electrical axis deviation and other cardiac diseases.
Diagnosis The range of filtering in mode of diagnosis is 0.05hz-130hz Monitoring The range of filtering in mode of monitoring is 0.5hz-40hz HARDEST The range of filtering in mode of HARDEST is 5hz-20hz Operation The range of filtering in mode of
SpO2 measurement
Real time measurement of SpO2, heart rate and perfusion index(PI).
Real-time transmission of pulse wave signals based on infrared spectrum absorption.
Real-time transmission of module’s state on working, hardware, software and sensor.
The host can issue an alarm according to relevant information.
Three patient models are supported: adult, pediatric and neonate.
Average time setup for parameters calculation to obtain different response times of parameter calculations.
The ability of anti-motion interference and weak signal measurement.
Manual Once: measurement under one command
STAT: Auto-measurement at regular intervals, 1min, 2min, 3min, 4min, 5min, 10min, 15min, 30min, 60min, 90min, 2h, 3h, 4h, 8h are for selection. The periodic time is the time interval between two measurements. If the interval from measurement ending to next measurement starting is less than 30 seconds, it will be automatically extended to 30 seconds
Continuous: Continuous measurement lasts 5 minutes, with a 5s interval between two measurements
NIBP measurement
The systolic pressure (SYS), mean pressure (MAP), diastolic pressure (DIA) and heart rate (PR) are included in measurement results.
Three measurement modes are provided: manual, automatic (STAT) and continuous.
Three patient models are supported: adult, pediatric and neonate.
Pressure calibration: continuous cuff pressure is provided to calibrate accuracy.
Leakage test: detect the integrity of airway.
Anti-motion interference and weak signal measurement.
High reliability and accuracy.
Real-time transmission of module’s state on working, hardware, software and sensor. The host can issue alarms according to relevant information.
Double overvoltage protection (hardware and software).
Double time-out protection (module time-out protection, providing timing trigger port for host).
Performances
ECG
Range of measurement
Range: 0 – 100 %
Base resistance value: Undefined
Variable resistance value: 1%
Range of PR
Range: .15-300 bpm (adult) / 15-350 bpm (pediatric/neonate)
Accuracy: ±1bpm
Resolution: . 1bpm
Noise: < 30 uv(peak-to-valley)
Input impedanc: .> 5 MΩ
Common-mode rejection
Diagnosis: . ≥90 dB
Monitoring: .≥100 dB
Operation: . ≥100 dB
Voltage range of electrode polarization: .±400mV
Baseline recovery time: < 3s after defibrillation
Detection for pacer signal
Detection amplitude: ±2 ~±700 mV
Pulse width: standard requirement–(0.1 ms-2.0ms)
Rise Time: 10us-100us
Electrotome protection
Cut mode: 300 W
Noise suppression for electrotome: Under the guidance of IEC 60601-2-27, article 5.2.8.14, the peak-to-valley noise tested with standard ECG lead wire is less than or equal to 2mV compared with the baseline
Mean algorithm of HR: Remove the maximum and minimum values of recent 12 RR intervals, and calculate the
mean value
Response to irregular rhythm
Figure 3a (Ventricular Bigeminy): 80±1 bpm
Figure 3b (Slow Alternating Ventricular Bigeminy): 60±1 bpm
Figure 3c (Rapid Alternating Ventricular Bigeminy): 120±1 bpm
Figure 3d (Bi-directional Systoles): 90±2 bpm
Response time of change in heart rate
Conform to requirement: IEC 60601-2-27
Heart rate from 80-120bpm: no more than 7s
Heart rate from 80-40bpm.: no more than 7s
Time to alarm for tachycardia: no more than 10s
Tall T-wave rejection capability: The maximum T-wave amplitude (aT) meet the requirement of 1.2mv specified in IEC
60601-2-27PVCS (statistics of premature beat)
Range: 0-300
Accuracy: No
Resolution: ±1
Analysis of ST-segment
Measurement range: – 2.0~2.0 mV
Accuracy: within the range of -0.8-0.8 mV,±0.02 mV or 10 %,
choose the bigger one; other ranges are not definedUpdate time: .10s
Respiration
Range of RR: 0-120 rpm
Accuracy: .15-120 rpm: ±2 rpm or ±2 %;
Base impedance: 500-2000 Ω
Variable impedance: 0.2-3.0 Ω
Body temperature
Range of measurement: 0-50°C
Accuracy: ±0.1C
Resolution: 0.1°C
SpO2
Range of measurement: 0-100%
Accuracy: 70-100%, ±2%;
Resolution: 1%
Pulse Rate (PR)
Range of measurement: 25–250bpm
Accuracy: ±2%
Resolution: 1bpm
Perfusion Index (PI)
Range of measurement: 0–20 %
Accuracy: No definition
Resolution: 0.001%
Cuff pressure(PI)
Range of measurement: . 0–300 mmHg
Accuracy: ±2mmHg or ±1% of reading
Resolution: 1mmHg
Blood pressure and pulse rate
Patient
categorySYS MAP DIA PR Adult 40–270mmHg 20–230mmHg 0 ~ 100% 30–240bpm Pediatric 40–200mmHg 20–175mmHg From 70 ~ 100%, ±2%;
not specified below 70%30–240bpm Neonate 40–130mmHg 20–100mmHg 0.025% 40–240bpm Measurement parameter Accuracy Resolution Blood pressure value The mean deviation <±5mmHg
The standard deviation <8mmHg1mmHg Measurement Accuracy ±2bpm or ±2% of reading
(take the bigger one)1bpm OVERPRESSURE PROTECTION
Patient category Protection pressure
of softwareProtection pressure
of hardwareAdult <297±3 mmHg <320±5mmHg Pediatric <250±3 mmHg <260±5mmHg Neonate <150±3 mmHg <160±5mmHg Maximum limit time for blood pressure measurement
Adult: < 120s
Pediatric: < 120s
Neonate: < 90s