Excellent Performance:
• LowPulseStr TM SPO₂ algorithm more reliable readings of SPO₂ during low perfusion and motion.
• iFastBP® NIBP algorithm used for fast and comfortable measurement.
• wSmartHeart ECG technology for more safey monitoring of patient.
• iRealRespTM Breating rate technology to get the real reading during motion.
• wSmartGas Capnograph technology to get reliable reading during multi-environment.
Powerful function:
• Multi display mode other beds view/Big font.
• 10,12,15 inch screen can be optional.
• Wi-Fi connectivity more flexible application.
• eTeleViewTM Central Montior system support.
• Easy to use with excellent usability.
Multi-scenario application: Bedside care/General care
CENTRAL MONITORING SYSTEM
• Central Monitoring System supports up to 64 beds or 64 patients across clinical units at the same time.
• 240 hours of 64-channel holographic physiological waveform storage and review.
• Provides review of up to 240 hours trend data storage, 720 alarm events per beds.
• Bi-directional communication with Aramis Series monitors for enhanced patient care.
| Standard application | Optional application |
| 3/5leads ECG | 12Leads ECG |
| NIBP | IBP*2 |
| RESP | C.O. |
| SPO₂ | EtCO2 (Main stream , TiniStream) |
| PR | Central monitor |
| HR | Touch Screen |
| TEMP | Recorder |
GENERAL
| Aramis 10: | Aramis 12: | Aramis 15: |
| Dimensions: 47*35*24cm | Dimensions: 47*35*24cm | Dimensions: 57*35*24cm |
| Weight: 3.8kg | Weight: 4.6kg | Weight: 5.4kg |
| Screen size: 10 inch | Screen size: 12 inch | Screen size: 15 inch |
PARAMETER
| SpO₂ | ECG |
| SpO2 Range: 0~100% | ECG Range: 0.15~5.5mV |
| SpO2 Accuracy: 70~100%, ±2% <70%, Undefined | ECG Resolution: 2.36uV/LSB |
| PI Range: 0~20% | HR Range: Adult: 15~300bpm Pediatric, Neonate: 15~350bpm |
| PVI Range: 0.001% | HR Accuracy: ±1bpm or ±1% (whichone is greater) |
| PR Range: 25~300bpm | RR Range: Adult: 0~120rpm Pediatric, Neonate: 0~150rpm |
| PR Accuracy: ±3bpm | RR Accuracy: 15~150rpm: ±2rpm or ±2% <15rpm: Undefined |
TEMP
Range: 0-50°C
Accuracy: ±0.1°C
Resolution: 0.1°C
RESP
Range: 0-120rpm
Accuracy: 15-120rpm; ±2rpm or ±2% of the reading (which one is larger); Others, undefined
Resolution: 1rpm
NIBP
Static Pressure Range: 0~300mmHg
Static Pressure Accuracy: ±2mmHg or ±1% of reading (take the larger value)
Static Pressure Resolution: 1mmHg
Blood pressure SYS Range: Adult: 40-270mmHg
Pediatric: 40-235mmHg
Neonate: 25-130mmHg
Blood pressure DIA Range: Adult: 10-210mmHg
Pediatric: 10-200mmHg
Neonate: 10-90mmHg
Blood pressure Mean Range: Adult: 20-230mmHg
Pediatric: 20-225mmHg
Neonate: 15-100mmHg
Blood pressure Accuracy: The mean deviation<±5mmHg The standard deviation<8mmHg
CO (Optional)
Range: 0.20~20.00L/Min
Accuracy: ±0.2 L/Min or ±5% of the reading (which one is larger)
IBP (Optional)
Pressure Range: -50~400mmHg
Accuracy: ±2mmHg or ±1%
PR Range: 20~350bpm
PR Accuracy: ±3bpm
EtCO2 (Optional)
CO₂ Range: 0~20.0vol%
CO₂ Accuracy: 0~12vol%: ± (0.2vol% + 2% of reading)
12~20vol%: ± (0.2vol% + 6% of reading)
AwRR Range: 0~150rpm
AwRR Accuracy: mainstream: 0~150rpm, ±1rpm
Sidestream: 0~70rpm, ±1rpm
71~150rpm, Undefined