Excellent Performance:
• LowPulseStrTM SpO2 algorithm provides more reliable SpO2 readings during low perfusion and exercise ;
• iFastBP® NIBP algorithm for fast and comfortable measurements ;
• wSmartHeart® ECG technology for safer animal monitoring ;
• wSmartGas® Capnograph technology for reliable readings in multiple environments .
Powerful function:
• Multi display mode other beds view/Big font ;
• 10,12,15inch screen can be optional ;
• Wifi connectivity more flexible application ;
• Smart alarm system more safety and more efficient ;
• Fanless design .
Multi-scenario application:
| Standard application | Optional application |
| 3/5leads ECG | C.O. |
| NIBP | IBP*2 |
| RESP | MainStream EtCO₂ |
| SpO₂ | TiniStream® EtCO₂ |
| PR | CMS |
| HR | Touch screen |
| TEMP | Printer |
| Wi-Fi |
PARAMETER
| SpO₂ | ECG |
| SpO₂ Range: 0~100% | ECG Range: 0.15~5.5mV |
| SpO₂ Accuracy : 70~100%,±2% <70%,Undefined | ECG Resolution: 2.36uV/LSB |
| PI Range: 0~20% | HR Range : 15~350bpm |
| PVI Range: 0-100% | HR Accuracy : ±1bpm or ±1% (whichever is greater) |
| PR Range: 25~350bpm | RR Range : 0~120bpm |
| PR Accuracy: ±3bpm | RR Accuracy : 6~120rpm : ±2rpm or ±2% <6rpm:Undefined |
TEMP
• Range: 0-50°C
• Accuracy: ±0.1°C
• Resolution: 0.1°C
NIBP
• Pressure Range : 0~300mmHg
• Pressure Accuracy : ±2mmHg or ±1% of reading (take the larger value)
• Resolution : 1mmHg
• SYS Range : Large animal (40-270mmHg), Medium animal (40-235mmHg), Small animal (25-130mmHg)
• DIA Range : Large animal (10-210mmHg), Medium animal (10-200mmHg), Small animal (10-90mmHg)
• Mean Range : Large animal (20-230mmHg), Medium animal (20-225mmHg), Small animal (15-100mmHg)
• Accuracy : The mean deviation < ±5mmHg The standard deviation < 8mmHg
IBP (Optional)
• Pressure Range : -50~400mmHg
• Accuracy : ±2mmHg or ±1% of reading (take thelarger value)
• PR Range : 20~350bpm
• PR Accuracy : ±3bpm or ±2% (take thelarger value)
C.O. (Optional)
• Range : 0.20~20.00L/Min
• Accuracy : ±0.2L/Min or ±5% (take thelarger value)
EtCO₂ (Optional)
• CO₂ Range: 0~20.0vol%
• CO₂ Accuracy: 0~12vol%: ± (0.2vol% + 2% of reading) 12~20vol%: ± (0.2vol% + 6% of reading)
• AwRR Range: 0~150rpm
• AwRR Accuracy: MainStream : 0~150rpm, ±1rpm, SideStream : 0~70rpm, ±1rpm, >70rpm, Undefined