Dedicated veterinary algorithm:
• LowPulseStrTM SPO₂ algorithm more reliable readingsof SPO₂ during low perfusion and motion.
• iFastBP® NIBP algorithm used for fast and comfortable measurement.
• wSmartHeart® ECG technology for more safey monitoring of patient.
• wSmartGas® Capnograph technology to get reliable reading during multi-environment.
Powerful function:
• Plug-in module, multiple parameters available.
• 12.1,15.6 inch screen can be optional.
• Wifi connectivity more flexible application.
• eTeleViewTM Central Montior system support.
• Smart alarm system more safety and more efficient.
• Fanless design.
Multi-scenario application:
| Standard application | Optional application |
| 3/5leads ECG | 12Leads ECG |
| NIBP | IBP |
| RESP | CO |
| SpO2 | EtCO2(Main stream , Side stream , TiniStream), |
| PR | Central monitor |
| HR | Touch screen |
| TEMP | Recorder |
PARAMETER
| SpO₂ | |
| SpO₂ Range: 0~100% | |
| SpO₂ Accuracy : 70~100%,±2% <70%,Undefined | |
| PI Range: 0~20% | |
| PVI Range: 0.001% | |
| PR Range: 25~350bpm | |
PR Accuracy: ±3bpm |
| ECG | |
| ECG Range: 0.15~5.5mV | |
| ECG Resolution: 2.36uV/LSB | |
| HR Range : 15~350bpm | |
| HR Accuracy : ±1bpm or ±1% (whichever is greater) | |
| RR Range : 0~120bpm | |
| RR Accuracy : 5~120rpm : ±2rpm or ±2% <5rpm:Undefined |
NIBP
• Pressure Range : 0~300mmHg
• Pressure Accuracy : ±2mmHg or ±1% of reading (take the larger value)
• Resolution : 1mmHg
• SYS Range : Large animal (40-270mmHg), Medium animal (40-235mmHg), Small animal (25-130mmHg)
• DIA Range : Large animal (10-210mmHg), Medium animal (10-180mmHg), Small animal (10-90mmHg)
• Mean Range : Large animal (20-230mmHg), Medium animal (20-200mmHg), Small animal (15-100mmHg)
• Accuracy : The mean deviation < ±5mmHg The standard deviation < 8mmHg
TEMP
• Range : 0-50°C
• Accuracy : ±0.1°C
• Resolution : 0.1°C
RESP
• Range : 0-120Rpm
• Accuracy : 15-120Rpm ±2Rpm or ±2% of the reading (which one is larger) Others,undefined
• Resolution : 1Rpm
IBP(Optional)
• Pressure Range : -50~400mmHg
• Accuracy : ±3mmHg or ±1% of reading
• PR Range : 25-250bpm
• PR Accuracy : ±2bpm or ±2% (whichone is greater)
CO(Optional)
• Range : 0.20~20.00L/Min
• Accuracy : ±5%
• AwRR Accuracy : mainstream : 0~150rpm,±1rpm Sidestream:0~69rpm,±1rpm 70~150rpm,Undefined