Excellent performance:
• LowPulseStr™ SpO₂ algorithm more reliable readingsof SpO₂ during low perfusion and motion.
• iFastBP® NIBP algorithm used for fast and comfortable measurement.
• wSmartHeart® ECG technology forsafer animal monitoring.
• iRealResp™ Breating rate technology to get the real reading during motion.
• wSmartGas® Capnograph technology to get reliable reading during multi-environment.
Powerful function
• Multi display mode Big font.
• 8 inch screen can be optional.
• Wi-Fi connectivity more flexible application.
• eTeleViewTM Central Montiorsystem support.
• Smart alarm system more safety and more efficient .
• Fanless design.
• Compact design and easy to use.
Multi-scenario application:
PMVS VET
Standard application SpO₂
NIBP PR
TEMP Optional application 3 Leads ECG
5 Leads ECG
EtCO₂ (Mainstream, TiniStream) Wi-Fi
Printer
Touch Screen
GENERAL
Dimensions: 17.5cm*27cm*12.5cm Weight: 2.4kg (Including battery, built-in module) Screen size: 8 inches
PARAMETER
SpO₂Range: 0~100% SpO₂ Accuracy : 70~100%,±2% <70%,Undefined PI Range : 0~20% PVI Range : 0-100% PR Range : 25-350bpm PR Accuracy : ±3bpm
ECG
| ECG Range : 0.15~5.5mV | |
| ECG Resolution : 2.36uV/LSB | |
| HR Range : 15~350bpm | |
| HR Accuracy : ±1bpm or ±1% (whichone is greater) | |
| RR Range : 0~120bpm | |
| RR Accuracy : 6~120rpm : ±2rpm or ±2% <6rpm:Undefined |
ETCP
Range 0-50°C Accuracy ±0.1°C Resolution 0.1°C
NIBP
| Pressure Range | 0~300mmHg |
| Pressure Accuracy | ±2mmHg or ±1% of reading (take the larger value) |
| Resolution | 1mmHg |
| SYS Range | Large animal : 40-270mmHg Medium animal : 40-235mmHg Small animal : 25-130mmHg |
| DIA Range | Large animal : 10-210mmHg Medium animal : 10-200mmHg Small animal : 10-90mmHg |
| Mean Range | Large animal : 20-230mmHg Medium animal : 20-225mmHg Small animal : 15-100mmHg |
| Accuracy | The mean deviation<±5mmHg The standard deviation<8mmHg |
EtCO₂ (Optional)
| CO₂ Range :0~20.0vol% | |
| CO₂ Accuracy :0~12vol%:±(0.2vol%+2%of reading) 12.1~20vol%:±(0.2vol%+6%of reading) | |
| AwRR Range :0~150rpm | |
| AwRR Accuracy :Mainstream:0~150rpm,±1rpm Sidestream:0~70rpm,±1rpm >70rpm,Undefined |